Ravaye E2

Vein Finder — near-infrared vein visualisation system. Version Poly2025.

The Ravaye™ E2 uses near-infrared light to detect veins beneath the skin, then projects their position onto the skin surface directly above them. Qualified medical personnel can observe the displayed vasculature to assist in locating a vein of suitable size and position for venipuncture and other procedures requiring the location of superficial veins.

Held directly overhead, the device accurately locates the centre line of a vein. Increasing displacement from directly overhead results in an offset in the displayed position, so the centre line — not the displayed width — should always be used as the target.

  • Non-invasive
  • Contact-free
  • Portable handheld
  • DLP projection
  • Class I
Ravaye E2 Vein Finder, a handheld near-infrared vein visualisation device
Ravaye™ E2 Vein Finder — handheld unit

Section 01

Intended use

The Ravaye™ E2 Vein Finder is intended to assist qualified healthcare professionals in visualising certain superficial veins. The device uses near-infrared illumination to enhance the visibility of subcutaneous veins, projecting their position directly onto the skin surface.

It is intended to be used as a supplement to appropriate medical training and experience. It is not a substitute for clinical judgment and should not be used as the sole method for locating veins. Practitioners should always follow the medical protocols of their facility.

Important. The device does not replace clinical judgment, standard venipuncture practice, or confirmation of vessel suitability. It is not a diagnostic device and is not intended for treatment of any kind.

Section 08

Technical specifications

From the Ravaye™ E2 user manual, document reference PEL-VF-100-UM-US Rev. 1.0.

Detection method
Near-infrared illumination technology
Infrared wavelength
760–940 nm
Detection depth
0–10 mm (0–0.39 in)
Optimal working distance
20–26 cm (7.9–10.2 in)
Vessel position accuracy
±0.3 mm (±0.012 in)
Vessel resolution accuracy
±0.3 mm (±0.012 in)
Projection technology
DLP projection
Working noise
≤ 35 dB
Battery
Rechargeable lithium, 4,900 mAh
Battery duration
Approximately 3.5 hours
Power supply
5 V / 4.5 A · Input 100–240 V, 50/60 Hz
Weight
450 g (15.9 oz)
Dimensions (L×W×H)
24 × 6 × 6.5 cm (9.4 × 2.4 × 2.6 in)
EMC
IEC 60601-1-2 · CISPR 11 Group 1, Class B

Specifications are subject to change in the interest of product improvement. Refer to the device labelling for the configuration of a specific unit.

Section 06

Display settings

The projection is adjustable, because vasculature, skin and room lighting are not consistent between patients.

12 options

Colour

White, yellow, purple, red, indigo, green, blue, and five paired combinations.

9 modes

Mode

Default, enhance, hair removal, guide line, and five mixture colours.

3 sizes · 6 levels

Size & brightness

Projection size L, M or S; brightness across six levels.

USB-C

Photograph

Save real-time vein images, retrievable over USB Type-C.

In practice. Default mode is used to scan a region quickly. Once a candidate vein is found, enhance mode improves visibility of deeper vasculature.

Section 05

Use and operation

  1. Position the device

    Hold 20 to 26 cm above the skin surface and scan the area of interest.

  2. Centre over the vein

    Ensure the projected display is centred directly above the vein’s centre line. Tilting offsets the projection from the true position beneath the skin.

  3. Optimise the image

    Adjust height and angle slightly. Moving closer or further can bring additional veins into view depending on vasculature, room lighting and vein depth.

  4. Scan, then enhance

    Use default mode to narrow down locations, then switch to enhance mode for deeper veins.

  5. Confirm before the procedure

    Verify location and suitability using standard medical techniques — visualisation, palpation and sound clinical judgment.

Section 02 & 10

Known limitations

Stated plainly, because a procurement team will find them anyway and would rather read them here.

The device displays only superficial vasculature and does not indicate vein depth. The maximum depth at which veins are displayed varies by patient.

Display quality is affected by vein depth, skin conditions such as eczema or tattoos, hair, scarring, highly contoured skin surfaces and adipose tissue. Some patients’ veins, or a portion of them, may not display well or at all. Projection may not display veins under bright light such as direct sunlight.

The device must not be used to locate veins in the eyes, and the display light should not be shone into the eyes for a prolonged period.

Section 03 & 07

Supplied with the device

Main unit
Ravaye™ E2 Vein Finder
Charging
USB Type-C cable and adapter
Case
Aluminium carrying case
Documentation
User manual, product certificate, warranty card
Calibration
Calibration card
Optional
Table stand and mobile stand, available separately

Projection accuracy is verified against the supplied calibration card before first use and periodically according to facility procedures. A device that fails calibration should be withdrawn from clinical use and serviced.

Section 13

Regulatory and manufacturing

Classification
Class I Medical Device · 510(k)-Exempt · Product Code KZA
United States
U.S. FDA Registered Establishment & Device Listed
Registration No. 3027571036 · Listing No. D588842
India
CDSCO Class A Medical Device · Registration No. POLYMA-Kanch-TN/M/IVD/021765
Quality systems
ISO 13485:2016 · ISO 9001:2015 · ISO 14001:2015
Europe
CE Marked under the Low Voltage Directive 2014/35/EU
Manufacturer
Polymatech Electronics Limited
SIPCOT Hi-Tech SEZ, Plot No. OZ 13, Oragadam, Kancheepuram District, Greater Chennai, Tamil Nadu 602105, India
U.S. Agent (FDA)
Nisene Technology Group
417A Salinas Road, Royal Oaks, CA 95076, USA
Document reference
PEL-VF-100-UM-US · Rev. 1.0 · July 2026
Registrations, availability and applicable editions differ by market. The specifications above are drawn from the United States edition of the user manual. Contact us for the edition and registration status applicable to your territory.

The name

Named for the Sanskrit Ravi — the Sun, the source of light and life.

Ravaye™ is a trademark of Polymatech Electronics Limited.